Search In this Thesis
   Search In this Thesis  
العنوان
Dosimetric verification of the performance of planning software in cases using different treatment techniques nasopharyngeal carcinoma /
الناشر
Mohamed Saber Ibrahim Elbasuny ,
المؤلف
Mohamed Saber Ibrahim Elbasuny
هيئة الاعداد
باحث / Mohamed Saber Ibrahim Elbasuny
مشرف / Wael M Elshemey
مشرف / Ehab M Attalla
مشرف / Mostafa Elnagar
تاريخ النشر
2019
عدد الصفحات
76 P. :
اللغة
الإنجليزية
الدرجة
الدكتوراه
التخصص
الإشعاع
تاريخ الإجازة
27/7/2019
مكان الإجازة
جامعة القاهرة - كلية العلوم - Medical Biophysics
الفهرس
Only 14 pages are availabe for public view

from 101

from 101

Abstract

The majority of developing countries are suffering from the lack of advanced radiotherapy treatment techniques. This study aimed at evaluating the performance of relatively available technique in developing countries (Three- Dimensional Conformal Radiotherapy, 3D-CRT) compared to Intensity Modulated Radiotherapy (IMRT) in the treatment of different stages of Nasopharyngeal Carcinoma (NPC). Forty NPC patients were selected and divided into two equal groups according to their diagnostic stages. The dosimetric performance of three different planning techniques; 3D-CRT, 7-field IMRT (7F-IMRT) and 9-field IMRT (9F-IMRT) were compared. The dose prescriptions were 70 and 66 delivered in 35 fractions to the sites of gross Planning Target Volume (PTV1) and bilateral retropharyngeal carcinoma (PTV2), respectively. Dose data for all of the three investigated planning techniques obeyed the Radiation Therapy Oncology group (RTOG) protocol no. 0615 (with few exceptions in 3D-CRT of Organs at Risk, OARs)